DAWN VIR CALIBRATION DOCUMENT

Version 2.9 (December 2017)

Written by: G. FILACCHIONE (INAF-IAPS, Rome), E. AMMANNITO (INAF-IAPS, Rome)

Document Change History

   ===========================================================================
   Change      Date          Affected Portions          Checked by   Approved by
   ===========================================================================
   
   Version 1.0 January 2011                                           A. CORADINI
                                                                      (INAF-IAPS, 
                                                                      Rome)
   
   Version 2.0 October 2013  Chapter 6 : added details  M. GIARDINO   M.C. DE SANCTIS
                             about the responsivity     (INAF-IAPS,   (INAF-IAPS, 
                             formula.                   Rome)         Rome)
                             Chapter 7 : added some                   M.T. CAPRIA
                             explanation about usage                  (INAF-IAPS, 
                             of internal calibration                  Rome)
                             Chapter 8 : added details 
                             about the dark current 
                             subtraction procedure,
                             corrected CALIB directory 
                             file names.
                             
   Version 2.1 November 2013                            M. GIARDINO   M.C. DE SANCTIS
                                                        (INAF-IAPS,   (INAF-IAPS, 
                                                        Rome)         Rome)
                                                                      M.T. CAPRIA
                                                                      (INAF-IAPS, 
                                                                      Rome)

   Version 2.4 May 2014       Chapter 8: added the      M. GIARDINO   M.C. DE SANCTIS
                              formula to calculate      (INAF-IAPS,   (INAF-IAPS, Rome)
                              the calibrated            Rome)         Rome)
                              reflectance factor                      M.T. CAPRIA 
                                                                      (INAF-IAPS, Rome)
                                                                                   
   Version 2.5 January 2016   Chapter 8: added the      M. GIARDINO   M.C. DE SANCTIS
                              name of the PDS field     (INAF-IAPS,   (INAF-IAPS, Rome)
                              containing the            Rome)         M.T. CAPRIA
                              integration time;                       (INAF-IAPS, Rome)
                              added information 
                              about the ITF file 
                              format.        
                                                                                   
   Version 2.6 October 2016   Chapter 4: changed        M. GIARDINO   M.C. DE SANCTIS
                              denomination from         (INAF-IAPS,   (INAF-IAPS, Rome)
                              spectral shift to         Rome)
                              spatial shift
                              Chapetr 8: added 
                              description of the 
                              detilt algorithm for 
                              the visible channel 
                                                                                   
   Version 2.7 February 2017  Chapter 8: added          E. AMMANNITO   M.C. DE SANCTIS
                              explanataion about the    M. GIARDINO   (INAF-IAPS, Rome)
                              parameters used in the    (INAF-IAPS,
                              detilt algorithm;         Rome)
                              updated the spectral 
                              and temporal range 
                              affected by the ITF 
                              null values 
                                                                                   
   Version 2.8 September 2017 Chapter 7: added          E. AMMANNITO   M.C. DE SANCTIS
                              description of internal   M. GIARDINO    (INAF-IAPS, Rome)
                              calibration procedure.    (INAF-IAPS, 
                              Created the new chapter   Rome)
                              8 containing the 
                              description of the 
                              calibration refinement 
                              obtained with the last 
                              ITF. Previous chapter 8
                              ("how to calibrate vir 
                              in-flight data") has 
                              been changed to chapter 
                              9, while previous 
                              chapter 9("references") 
                              has been changed to 
                              chapter 11.
                              Chapter 9: added an 
                              explanation about the 
                              presence of 
                              contamination across 
                              campaigns. Created 
                              chapter 10 containing 
                              the listing of knwon 
                              instrument artifacts and
                              their effects on data 
                              usability
                              Chapter 11: added a 
                              new reference  
                                                                                   
   Version 2.9 December 2017  General revision          E. AMMANNITO   M.C. DE SANCTIS
                              applied, during the PDS   M. GIARDINO    (INAF-IAPS, Rome)
                              peer review performed     (INAF-IAPS, 
                              on the Ceres volumes      Rome)
                              between July and 
                              December 2017
                              The internal 
                              calibration steps have 
                              been further documented 
                              by inserting a table in 
                              Chapter 7.
                              In Chapter 9, the 
                              change operated to the 
                              execution cycle of the 
                              instrument to avoid the 
                              contamination found in 
                              VSH and VH2 is 
                              described. 
   ===========================================================================  

1. INTRODUCTION

This document describes the algorithms used to calibrate VIR raw (EDR) data to data in physical units (RDR, spectral radiance), in order to give the final user a detailed view of the methods used to remove instrumental effects on the data. A complete calibration campaign of VIR was performed at the channel level in Selex Galileo (SG), Florence by using a variety of calibration facility setups. These calibrations were performed immediately after instrument assembly and before delivery to Orbital for integration with the Dawn spacecraft. The SG calibration included spectral, geometrical, flat-field and radiometric measurements as well as characterization of the detectors' performances (including defective pixels, linearity and dark current at various operative temperatures), the instrumental thermomechanical stability, the data-commanding-telemetry handling, and electrical interfaces. A description of the methods used and results of these tests are described in De Sanctis et al. (2011).

Section 2 describes the experimental setup used for pre-launch calibrations at SG; section 3 is devoted to the description of the spectral calibration; geometrical calibration is included in section 4; flat-field is described in section 5; section 6 contains a description of radiometric calibration; section 7 is about the internal calibration procedure; section 8 describes an in-flight instrument response function update using the onboard lamp to improve the calibration in the spectral range 2.5 µm - 3.5 µm; and finally section 9 explains the algorithms used to convert raw data into spectral radiance or reflectance.

2. CALIBRATION SETUP

The basic setup used during the calibrations consists of an optical bench over which are housed a collimator, a reference target placed at its focal plane and a folding mirror used to move the collimated beam in the instrumental FOV along the azimuthal (sample) and zenithal (line) directions. Since VIR focuses at infinite distance it becomes necessary to use a collimator to have a collimated reference beam impinging the optical pupil. The SG-developed collimator uses an off-axis parabola (D=250 mm, F=1020 mm, off axis angle=8°), which guarantees an unobstructed beam, reduced aberrations, and high spatial scale. For VIR the magnification ratio is equal to:

MR=F_VIR/F_collimator=152mm/1020mm=0.15

This means that 1 mm on the collimator's focal plane corresponds to 0.15 mm on the VIR focal plane. The VIR detector has a 40 µm pixel pitch (square), so this scale corresponds to 4 spatial pixels along both sample and line directions. The collimator's focal plane is equipped with a holder able to sustain several interchangeable targets (pinholes, test slits, MTF masks, and a matrix of 5x5 microlamps). These elements are used to perform the different calibrations. The collimated beam is folded towards the instrument by using a folding mirror placed over two computer controlled, micrometric mounts able to aim it with steps of 1 µrad along the azimuthal (scan parallel to VIR slit, along sample direction) and zenithal (scan perpendicular to the slit, along lines direction) angles.

During the calibration the VIR instrument is housed in a thermo-vacuum chamber in order to reproduce the operating conditions aboard the satellite. In these conditions it's possible to cool the IR detector down to the operating temperature of about 70 K by using the cryo-cooler (operating on a Stirling cycle) and the CCD to about 160 K by using a passive radiator. The collimated optical beam reaches the spectrometer's pupil through a CaF2 window housed in the front of the thermo-vacuum chamber. This window is characterized by an elevated optical transmittance in the 250-5100 nm spectral range.

All opto-mechanical devices placed on the optical bench are controlled by using a dedicated software (OCS, Optical Control System), while the VIR instrument is controlled by using a separate setup, consisting of the UT (Unit Tester) connected to the experiment through the Proximity Electronics Module (PEM). This system allows the sending of commands to the instrument, to start acquisitions only when all optical elements commanded by OCS are in the correct configuration, and to receive back and record telemetry and calibration data.

3. SPECTRAL CALIBRATION

The spectral calibration characterizes a fundamental aspect of the functional requirements of a hyperspectral imaging spectrometer: the conversion of bands positions along the spectral axis of the detectors into wavelength units. The spectral calibration is obtained through the following steps:

The following instrumental parameters are deduced from the spectral calibration:

Because the instrument uses a diffraction grating that disperses the light according to a linear law we can assume SSI(n) = SSI; in this case the spectral calibration relation assumes the following expression for the spectral dispersion of the central wavelength:

λc(n) = λ0 + SSI • b

These quantities were measured during the ground calibration by acquiring several fine spectral scans using a monochromator as a source. The calibration setups used to define the spectral properties of VIR used a heritage setup developed for the VIRTIS/M aboard Rosetta and the Venus Express missions (Ammannito et al. 2006; Filacchione et al. 2006). Two different configurations were used to characterize the spectral response, the first using a transmission method and the second using a diffusion method. In the first case (transmission) on the optical bench, the source, the monochromator, the test slit and the collimator were present; using this set-up the level of the signal was high enough to stimulate VIR, but the alignment between the output slit of the monochromator and the test slit of the optical bench was difficult to achieve. In the second case (diffusion) the source, the monochromator, a silvered diffusive target and the collimator were on the bench. In this case the alignment of the system wasn't critical, but the level of the signal was lower. The monochromator scans different wavelengths, thus illuminating the diffusive screen. VIR acquires monochromatic images at each step. Therefore VIR is "simulated" at wavelength steps smaller than its spectral resolution. In this way it is possible to associate to each frame the wavelength of the input beam coming from the monochromator. Studying the profile over the lines of each illuminated band, it is possible to get the spectral response function of that particular band.

3.1 Transmission method

The transmission method was applied only to the visual channel. For each illuminated pixel, the spectral response function was computed by fitting a Gaussian-like function over the profiles measured during the spectral scans. With this setup the intensity of the signal along the slit isn't uniform. This could be related to a misalignment between the output slit of the monochromator, the test slit of the optical bench in particular, and the entrance slit of the experiment. Moreover on the optical bench a spectral shift along the slit is apparent, given that at every illuminated band a different central wavelength is found. The central wavelength and the spectral width of the illuminated bands are calculated by averaging such parameters over every illuminated sample. The central wavelength of the band is calculated by using a linear fit while the spectral width is given by a polynomial fit. A summary of the parameters calculated with the fits and their uncertainties are reported in the following table:

           Spectral dispersion               Spectral width                           
                                                                                
Model       a • x + b                        a • x4 + b • x3 + c • x2 + d • x + e
Parameters  a=1.89297 b=245.744              a=5.25E–11 b=-6.08E–8 c=2.74E–5 d=-0.0049 e=2.13
Sigma       σa=0.00016 σb=0.041              σa=0.14E–11 σb=0.14E–7 
                                             σc=4.82E–5 σd=0.0068 σe=0.31
                                                                                

In the next table are indicated, for all the illuminated bands, the measured and calculated values of the central wavelength and spectral width. The measured values are averages computed over all the illuminated samples for a fixed band.

Band #              λmeas             λcal               SWmeas           SWcal
                     (nm)             (nm)                (nm)             (nm)
                                                                             
79                   395.125          395.289             1.89559          1.88451
                                                                             
80                   397.049          397.182             1.89332          1.88290
                                                                             
81                   398.964          399.075             1.88235          1.88131
                                                                             
82                   400.860          400.967             1.87189          1.87976
                                                                             
83                   402.761          402.860             1.86246          1.87824
                                                                             
84                   404.662          404.753             1.87010          1.87675
                                                                             
157                  542.828          542.94              1.87158          1.82947
                                                                             
158                  544.925          544.833             1.84376          1.82947
                                                                             
159                  546.835          546.726             1.83146          1.82949
                                                                             
160                  548.742          548.619             1.82906          1.82952
                                                                             
161                  550.640          550.512             1.82154          1.82956
                                                                             
162                  552.548          552.405             1.84028          1.82961
                                                                             
163                  554.454          554.298             1.84128          1.82967
                                                                             
237                  694.480          694.378             1.81275          1.85861
                                                                             
238                  696.323          696.271             1.86955          1.85922
                                                                             
239                  698.219          698.164             1.84252          1.85983
                                                                             
240                  700.101          700.057             1.85477          1.86045
                                                                             
241                  702.000          701.950             1.86519          1.86107
                                                                             
242                  703.898          703.843             1.85821          1.86170
                                                                             
317                  845.842          845.816             1.91248          1.91559
                                                                             
318                  847.74           847.709             1.91959          1.91639
                                                                             
319                  849.647          849.602             1.92232          1.91720
                                                                             
320                  851.542          851.495             1.91718          1.91800
                                                                            
321                  853.447          853.388             1.91865          1.91881
                                                                             
396                  995.282          995.361             1.97795          1.99122
                                                                             
397                  997.171          997.254             1.98837          1.99243
                                                                             
398                  999.063          999.147             1.99155          1.99365
                                                                             
399                 1000.95          1001.04              1.98800          1.99487
                                                                             
400                 1002.85          1002.93              2.00388          1.99611
                                                                             
401                 1004.75          1004.83              2.00954          1.99735
                                                                                

From the analysis of measurements we have noted the presence of a slight spectral shift occurring along the slit. In order to evaluate this effect we repeated the calculation of the linear fit coefficients for other samples along the slit (at samples = 110, 140); previously, the results at the slit's center (sample = 128) were discussed. This analysis demonstrates the presence of a change in the spectral calibration response when repeated on different points along the slit. The fit parameters with their errors on samples 110, 128, 140 are reported in the following table.

Sample#               λmeas              SWmeas
                      (nm)               (nm)                                   
                                                                                
110              246.76±0.56          1.8926±0.0022
                                                                                
128              245.83±0.30          1.8926±0.0011
                                                                                
140              245.40±0.19          1.8921±0.0007               

These calculations demonstrate that the parameters are mutually incompatible so the central wavelengths calculated with the coefficients indicated in the previous table cannot be used and another calibration approach is necessary. At the end of the post-processing analysis it was concluded that the measurements of the spectral width are satisfactory while further measurements are needed to determine the dispersion coefficients. Using the Transmission setup, in fact, they seem to be sample dependent. In the next paragraph the results obtained using the Diffusion setup data are discussed.

3.2 Diffusion method

The Diffusion method was used to characterize the spectral response of both the Visual and the Infrared channels. We use these measurements to determine the central wavelength of the VIS channel and both the central wavelength and the spectral width of the Infrared channel. Comparing these results with the similar profiles taken with the Transmission setup, it is apparent that the spectral shift among profiles taken at different samples is negligible. In this way it is verified that the effect is caused by the Transmission set-up characteristics (difficult to co-align VIR and test slit orientations) and is not due to the VIR malfunctioning.

Using the same technique discussed previously, we have retrieved the best spectral dispersion and width values. For both channels, the central wavelength of each band b is retrieved through a linear fit while for the spectral width a polynomial fit is used.

               Spectral dispersion          Spectral width
VIS            Model a • x + b              a • x4 + b • x3 + c • x2 + d • x + e
Parameters     a=1.89223 b=245.660          a=1.3E–10 b=-1.1E–7 c=1.89E–5 d=0.0047 e=1.6
Sigma          σa=0.00033 σb=0.085          σa=8.1E–10 σb=7.8E–7 
                                               σc=0.26–5 σd=0.0037 σe=1.7
                                                                                
IR             Model a • x + b              a• x4 + b • x3 + c • x2 + d • x + e
Parameters     a=9.4593 b=1011.29           a=-6.8E–10 b=8.23E–7 c=-2.09E–4 d=0.0021 e=13.9
Sigma          σa=0.0011 σb=0.28            σa=2.3E–9 σb=1.6E–6
                                               σc=3.80E–4 σd=0.0335 σe=1.2
                                                                                

We report in the next Table the measured and computed values of the central wavelength and spectral width for both channels. The measured values are averaged over all the samples for a given band.

VIS channel                                                                     
Band                λmeas             λcal               SWmeas           SWcal
#                   (nm)             (nm)                (nm)             (nm)
                                                                                
81                  398.875          398.931             2.07338          2.03109
                                                                                
82                  400.774          400.823             2.02554          2.03699
                                                                                
83                  402.597          402.715             1.99148          2.04287
                                                                                
84                  404.436          404.607             2.07068          2.04875
                                                                                
158                 544.700          544.633             2.42099          2.44303
                                                                                
159                 546.577          546.525             2.44812          2.44754
                                                                                
160                 548.465          548.417             2.45310          2.45203
                                                                                
161                 550.355          550.309             2.45829          2.45648
                                                                                
162                 552.249          552.201             2.47181          2.46091
                                                                                
163                 554.144          554.094             2.47652          2.46531
                                                                                
164                 556.028          555.986             2.46822          2.46969
                                                                                
238                 696.007          696.011             2.69859          2.70952
                                                                                
239                 697.892          697.903             2.69611          2.7116
                                                                                
240                 699.785          699.795             2.69541          2.71365
                                                                                
241                 701.677          701.688             2.70452          2.71567
                                                                                
242                 703.572          703.58              2.72082          2.71767
                                                                                
243                 705.508          705.472             2.80885          2.71963
                                                                                
317                 845.453          845.497             2.75994          2.79344
                                                                           
318                 847.337          847.389             2.80736          2.79368
                                                                         
319                 849.239          849.282             2.8061           2.79391
                                                                         
320                 851.142          851.174             2.79165          2.79412
                                                                         
321                 853.043          853.066             2.80487          2.79433
                                                                         
396                 994.789          994.983             2.75589          2.80388
                                                                         
397                 996.686          996.875             2.8257           2.80439
                                                                         
398                 998.595          998.768             2.77639          2.80492
                                                                         
399                 1000.46          1000.66             2.79735          2.80548
                                                                         
400                 1002.28          1002.55             2.78868          2.80606
                                                                         
401                 1004.22          1004.44             2.84745          2.80666
                                                                                
IR channel                                                                     
Band                λmeas             λcal               SWmeas           SWcal
#                   (nm)             (nm)                (nm)             (nm)
                                                                                
2                   1029.3           1030.21             14.0742          13.9467
                                                                        
3                   1038.77          1039.67             13.78            13.9478
                                                                         
103                 1986.31          1985.6              12.9869          12.7654
                                                                         
104                 1995.85          1995.06             12.7585          12.7477
                                                                         
105                 2005.35          2004.52             12.6494          12.7299
                                                                         
106                 2014.87          2013.98             12.5963          12.7122
                                                                         
208                 2978.82          2978.83             11.4667          11.4689
                                                                         
209                 2988.07          2988.29             11.2923          11.4666
                                                                         
210                 2997.45          2997.75             11.6996          11.4645
                                                                         
211                 3006.83          3007.21             11.5063          11.4626
                                                                         
212                 3016.13          3016.67             11.337           11.461 
                                                                         
315                 3991.6           3990.98             12.7906          12.9028
                                                                         
316                 4000.3           4000.44             12.6859          12.9335
                                                                         
317                 4010.2           4009.9              13.5031          12.9647
                                                                         
367                 4482.68          4482.86             14.9175          14.9333
                                                                         
368                 4492.2           4492.32             15.1484          14.9807
                                                                         
369                 4501.56          4501.78             14.8612          15.0283
                                                                         
370                 4511.02          4511.24             15.1136          15.0763
                                                                                

Following the post-processing analysis, we find that the measurements of the dispersion coefficients are compliant with the specifications for both the visual and infrared focal planes. For the VIS channel, computation with the diffusion method gives better results than computation with the transmission method. The quality of the spectral calibration was confirmed by observing the spectrum of a calibrated HgNe pencil lamp.

4. GEOMETRIC CALIBRATION

The geometrical calibration allows characterization of:

1. the field of view, hereafter FOV;

2. the instantaneous field of view (hereafter IFOV) of different pixels along and across the spectrometer's slit directions (respectively sample and line directions).

We define the pixel function, PF(s), as the convolution of a unitary step function V (s) (representing the real pixel) with the instrumental response along the sample direction, INST(s):

PF(s) = V(s) ⊗ INST(s)

The slit function, SF(l), is given by the convolution of a unitary step function U(l) (representing the spectrometer's slit response) with the telescope response along the line direction, TEL(l):

SF(l) = U(l) ⊗ TEL(l)

These two responses were measured during the pre-launch calibration campaign acquiring the signal produced by a test-slit, illuminated by a HgNe lamp, having an equivalent width narrower than the instrumental IFOV (the test slit aperture is 3.0 x 0.1 mm, corresponding to 12 x 0.4 pixels at VIR scale). This test-slit is placed at the collimator's focus and it is moved at subpixel steps perpendicular and parallel to the VIR slit by moving the folding mirror. By using this method it is possible to measure the FWHM of the IFOV at three positions of the FOV (boresight: sample = 128, line = 128), position N: sample = 38, line = 218; position O: sample = 218, line = 38). For the VIS channel the FWHM of the pixel function ranges over the 237.9–244.1 µrad interval while the slit function is 287.7–389.4 µrad; for the IR channel the ranges are 421.7–488.1 and 350.9–367.3 µrad respectively. These differences are caused by a residual of astigmatism in the optical design.

The determination of the FOV (nominally 3.6° x 3.6°) is possible through the imaging of a 5 x 5 array of microlamps placed at the focus of a collimator. This array was built to cover the entire FOV when placed at collimator's focus: the presence of a regular grid of subpixel sources allows for evaluation of the imaging and geometrical performances of the experiment. The absolute position of each microlamp was measured with a theodolite placed on the pupil of the collimated beam; when compared to the relative positions of the lamps spots on the images it is possible to infer the dimensions of the instrumental FOV.

Moreover, this setup is particularly useful in evaluating the presence of possible "spatial shift", e.g. a mismatch between the position of one monochromatic image with respect to another. This effect is particularly evident on VIRTIS-M on Rosetta, where it reaches a shift of about 8 spatial pixel (samples) between the first and the last image of the VIS channel. The cause of it is a slight misalignment among slit, grating grooves and focal plane orientation (for a full discussion of the spectral tilt and post-processing corrective methods the reader can refer to Filacchione 2006). For VIR several optical improvements were made to the grating design that drastically reduce this effect. Analysis of the 25 microlamp target data allows verification that the spatial shift on the VIS channel reaches about 2 spatial pixels between the two spectral extremes of the range ( 255 nm and 1071 nm ). This value comes from the analysis of the distribution of the microlamps' position (in sample-line space) on the monochromatic images. As each microlamp has a subpixel dimension when seen by VIR through the optical bench setup, it is possible to measure the associated barycenter position through a 2D Gaussian fit; this procedure is done for each lamp and for every spectral band (432 images).

5. SPATIAL CALIBRATION: FLAT-FIELD

The flat-field is defined as the response of the instrument to a uniform source (Filacchione et al. 2006). It is used to homogenize the pixels' response across the whole focal plane. In the case of imaging spectrometers using 2D detectors, flat field matrices contain, for each wavelength, the relative variation of the instrumental response with respect to the boresight (sample s* = 127).

The measurements of the VIS and IR flat-field matrices were calculated during the pre-launch tests by acquiring a spatially flat source placed on the focus of a collimator and aligned to the VIR boresight. The source used in the 0.25–2.5 µm range is a Lambertian surface illuminated by a QTH lamp; this target is about 10 x 10 cm wide in order to completely fill the instrumental FOV. It is replaced by a blackbody source for the measurement of the flat-field in the 2.5–5.0 µm range. In both cases the flat field is retrieved through a spatial scan across these targets by moving the folding mirror at 1 IFOV step. This approach allows for observation of the same region of the target with each pixel (sample) of the detector, thus eliminating possible target non-uniformity from the flat-field matrices.

The resulting flat-field matrices for the two focal planes are given by the ratio of the signal measured at a certain position of the focal plane (b, s) with respect to the signal measured at boresight (s = s*) and at the same band position b:

FF(b,s) = Ns(b,s) / Ns(b, s*)

Flat-field matrices are sensitive to the characteristics of the detector (single and clusters of defective pixels, dis-uniformities due to the production process) and of the optical layout (the two horizontal features at samples 80 and 150 are caused by the slit's shape; several vertical features with a symmetry with respect to boresight are introduced by the grating design).

6. RADIOMETRIC CALIBRATION

As explained in section 5, the wide spectral range of the experiment can only be explored by using different sources (Filacchione et al., 2006). For the radiometric calibration two different sources are necessary:

The input radiance is measured and verified through a laboratory radiometer (a Field-Spec(TM) spectroradiometer). Unfortunately as the optical pupil of the Field-Spec optics does not match entirely with the VIR pupil, the measured radiance can only be used as a relative value. The absolute value of the radiance has been tested in flight during dedicated observations of known targets, such as stars (Arcturus and Canopus) and planets (Mars). The lamps used are observed first with the spectroradiometer and then with VIR. Knowing the value of input radiance, we can associate it with an average of 50 VIR acquisitions of the Spectralon(TM) target, taken at slit center, with an integration time ti = 10 s. The Responsivity, R, is therefore calculated by applying the following equation:

R(b, s*) = DN(b, s*) / (BB(b)*ti)

where R(b,s*) is the responsivity computed for each band b at the sample s*, DN(b,s*) is the raw signal in digital numbers acquired by the spectrometer for each band b at the sample s*, BB(b) is the radiance of the source measured by the reference spectroradiometer and sampled at the VIR spectral band b and ti is the integration time.

The expansion to the sample of the focal plane different from s* is possible applying the flat-field FF. In this way we retrieve the ITF (Instrument Transfer Function) array:

ITF(b, s) = FF(b, s) • R(b, s*)

The IR channel radiometric calibration is done by acquiring directly the radiance emitted by a blackbody source placed at the collimator's focus. The blackbody temperature is set at different values in order to have a good SNR on several spectral ranges and with different integration times (a summary of the acquisitions is given below). As reported in the next Table, only a limited spectral range can be evaluated for a given blackbody temperature and integration time: for bands < Min Band the signal is very low and it includes only the readout offset and residuals of the dark current; for bands > Max Band value the signal is saturated. Therefore, the responsivity is retrieved by using only the signal intervals as indicated in following:

TBB (°C)           ti (s)                Min Band               Max Band    
                                                                                
50                 0.2                   250                    438             
                                                                                
                   1.0                   238                    280             
                                                                                
                   2.0                   238                    255             
                                                                                
                   5.0                   170                    240             
                                                                                
100                0.2                   238                    281             
                                                                                
                   1.0                   148                    239             
                                                                                
                   2.0                   140                    195             
                                                                                
                   5.0                   120                    170             
                                                                                
200                0.2                   110                    174             
                                                                                
                   1.0                   80                     120             
                                                                                
                   2.0                   70                     105             
                                                                                
                   5.0                   65                     95              
                                                                                
300                0.2                   60                     100             
                                                                                
                   1.0                   40                     68              
                                                                                
                   2.0                   35                     58              
                                                                                
                   5.0                   0                      37              
                                                                                
350                0.2                   0                      78              
                                                                                
                   1.0                   0                      52              
                                                                                
                   2.0                   0                      35              
                                                                                

The IR responsivity is computed by using:

R(b, s*) = DN(b, s*) / (BB(b) * ti)

where the blackbody radiance BB is given by Planck's formula. Finally, applying

ITF(b, s) = FF(b, s) • R(b, s*)

to these data, it is possible to derive the responsivity for each pixel of the IR channel.

7. INTERNAL CALIBRATION

Instrumental performances were checked during in-flight conditions by using internal calibration sequences. VIR can acquire reference signals by using the combination of the cover, shutter and VIS and IR lamps (Melchiorri et al., 2003). These lamps, housed on the side of the telescope illuminate the internal side of the external cover. The cover is placed near the entrance pupil of the instrument to minimize optical aberrations. The window of each lamp contains a transparent filter (holmium for the VIS, polystyrene for the IR) to introduce some well-shaped spectral absorption features on the overall spectrum. The signal coming from the two lamps can be used to:

- check the in-flight stability of the instrument spectral response;

- check the in-flight stability of the flat-field;

- monitor the evolution of defective pixels (number and distribution);

- perform a check on the relative radiometric response of the instrument.

The internal calibration mode, implemented in the VIR on-board software, consists of the acquisition of a sequence of 35 frames organized into the following steps: 5 acquisitions of electronic offsets, 5 acquisitions of background signal, 5 acquisitions of dark currents, 5 acquisitions of the IR lamp signal, 5 acquisitions of the VIS lamp signal, 5 more acquisitions of dark currents and finally 5 acquisitions of background signal.

The electronic offset is measured by acquiring 5 frames with both detectors on with an exposure time of zero seconds, while maintaining the instrument cover closed and the shutter open, so to read only the "electronic noise produced inside" the instrument chassis, isolated from any signal coming from outside.

The background signal is instead measured by acquiring 5 frames in the same condition defined for the electronic offset, except for setting an exposure time greater than zero for both detectors.

The dark current frames are produced while keeping both the shutter and the cover closed and activating both detectors.

The measurements of the dark currents and that of the background signal are repeated twice during calibration mode, to record their values both before and after the activation of the internal calibration lamps. Finally, the signal from the calibration lamps is measured by activating the given calibration lamp (IR or VIS), with the shutter open and the cover closed with the internal side illuminated by the lamp itself.

Calibration is done illuminating with one lamp at a time.

In the following table these different phases are listed.

===========================================================================
Phase        Frames   Cover   Shutter   IR     VIS    Exp. Time   Exp Time
             Number                     lamp   lamp   VIS(s)      IR(s)
===========================================================================
Electronic   1-5      closed  open      off    off     0.0         0.0
Offset

Background   6-10     closed  open      off    off     1.0         0.5
1 

Dark         11-15    closed  closed    off    off     1.0         0.5
Current 1

IR Lamp      16-20    closed  open      on     off     20.0        0.5

VIS Lamp     21-25    closed  open      off    on      1.0         0.02

Dark         26-30    closed  closed    off    off     1.0         0.5
Current 2

Background   31-35    closed  open      off    off     5.0         5.0
2
===========================================================================  

[INTERNAL CALIBRATION SEQUENCE]

Data acquired during this sequence are fundamental to follow the temporal evolution of the instrument and to monitor the overall performances in operative conditions. They can also be used in the calibration pipeline.

8. CALIBRATION REFINEMENT

During the mission operations at Vesta, we identified some artifacts in the instrumental transfer function (ITF) in the 2.5–3.5 µm spectral range, where several absorption bands of OH and H2O occur. These artifacts are systematic and of the same relative magnitude in all images, and therefore they do not prevent the detection of relative spectral variations associated with OH and H2O on the surfaces of the target bodies. These artifacts are systematic errors due to the non-homogenous instrumental response of a detector producing nonphysical spectral signatures. The precise cause of the artifacts are not yet fully understood, but involve separately or in combination imperfect radiance calibration, spectral miscalibration, peculiar readout noise in the detector electronics, and from uncertainties in the solar reference spectrum. Nevertheless, we have devised a means to correct the ITF using the onboard lamps.

To compute the ITF for this range, we used in-flight data from one of the internal lamps of the spectrometer. This lamp, made of a tungsten filament, is characterized by a blackbody-like emission at about 2400 K. Since the spectrum of the infrared radiation emitted by these filaments is featureless, a polystyrene filter was inserted for a wavelength calibration of the IR channel. The blackbody radiation of the internal lamp has been used to retrieve a relative ITF in the 2.5–3.5 µm spectral range. First, we calibrated the signal from the internal calibration lamp with the on-ground response function, and then we retrieved the equivalent temperature of the radiation fitting a Planck curve, a value around 1500 Kelvin. The new ITF is the result of the ratio between the raw signal of the lamp and the Planck function. This ratio must be multiplied by a factor to take into account the integration time used to acquire the signal, the transmittance of the polystyrene filter, and the viewing geometry. We compared the ITF obtained with the on-ground calibration with the in-flight calibration and a combination of the two, and we analyzed the calibrated spectra computed with these three versions of the ITF: the comparison graphs between these three curves can be found in [8]. We observe that the new ITF minimizes most calibration residuals that were showing as artifact peaks between 2.5 and 2.9 µm in the previous calibration. The final ITF (version 2) is the ITF derived during the on-ground calibration campaign with the exception of the spectral channels between 2.5 and 3.5 µm where the ITF is derived using the method described here.

9. HOW TO CALIBRATE VIR IN-FLIGHT DATA

The VIR team receives data and telemetry packets from the satellite from the Dawn Science Center (UCLA-JPL). These packets are processed at the PI institution (INAF-IFSI, Rome, Italy) with proprietary GSE (Ground Support Equipment) and converted into standard PDS (Planetary Data System) format. Dedicated package scripts and routines and calibration files distributed with this archive are used to convert the raw data in physical units.

Only for the visible channel, a specific detilt algorithm must be applied to the raw cube to deal with the spatial shift described in chapter 4. This algorithm is applied as a first step in the calibration procedure, before any other step described in this section.

The spatial tilt effect in the visible channel can be compensated by a shift of equal magnitude directed towards the opposite axis of the sample.

The detilting is obtained through the following steps:

The value 20 as a multiplicative factor, has been chosen to find a suitable compromise between the accuracy of the algorithm and the computational time required to obtain the result.

The spatial offset applied is a constant value of 2 samples along the slit. As a result, the frame obtained has the same size of the original frame, but the two ending columns (corresponding to two samples) are filled with empty values and are unusable.

The routine implementing the steps above is shown in the following IDL code snippet: the input data for this routine is the raw cube (raw_qube), while the output data is the detilted raw qube(raw_qube_detilt).

frame_expanded=uintarr(bands,samples* 44)

frame_exp_detilt=uintarr(bands,samples * 40)

raw_qube_detilt=uintarr(bands,samples,lines)

; number of pixel by which each sample in each spectral channel has to be shifted 
for li=0,lines-1 do begin
     ; step 1: oversampling
     
     for sa=0,samples-1 do begin
     
          for ss=0,39 do begin
          
               frame_expanded(*,sa*40+ss)=reform(raw_qube(*,sa,li))
          
          endfor
          
     endfor
     
     
     
     ; step2: detilt
     
     for sa=0,samples-1 do begin
     
          for ba=0,bands-1 do begin
          
               bsh=ba/4
               
               for ss=0,39 do begin
               
                    frame_exp_detilt(ba,sa*39+ss)=frame_expanded(ba,sa*39+ss+bsh)
                    
               endfor
               
          endfor
          
     endfor
     
     
     ; step3: resampling to the original size
     
     for sa=0,samples-1 do begin
     
          tot=dblarr(bands)
          
          for ba=0,bands-1 do begin
          
               for ss=0,39 do begin
               
                    tot(ba)=frame_exp_detilt(ba,sa*40+ss)+tot(ba)
                    
               endfor
               
               raw_qube_detilt(ba,sa,li)=tot(ba)/40.0
               
          endfor
          
     endfor
     
endfor                                                                          

The Software Interface Specification (SIS) document contains the details of VIR data cube format and processing steps, detailed briefly here. A raw data cube contains uncalibrated signal Ns in DN; dark currents are periodically stored in the same raw data cube and in each data cube there is at least one dark current acquisition. The dark current must be subtracted from the original data in the raw cube before the conversion in physical units. The number and location of dark current frames in each raw cube is documented in the hkt table (shutter status, open if normal acquisition, closed if dark current acquisition). The same information alternatively can be found by reading the parameter DARK_ACQUISITION_RATE in the data cube label file.

Raw data cubes may have one or more dark current frames. If there is only one dark current frame in a raw data cube, the equivalent dark current frame is the same for every frame in the data cube and is equal to the only dark current frame acquired, If there is more than one dark current frame, the equivalent dark current frame is the interpolation in time of two consecutively acquired dark frames. The dark subtracted frames are computed subtracting the equivalent dark current frame from the original frame in the raw data cube. At the end of this operation the dark current frames are removed, and there will be a dark subtracted data cube with the same bands and samples number of the raw data cube and a number of lines equal to the original minus the number of dark current frames. This is the reason why calibrated cubes have a lower number of frames than the corresponding raw cubes.

The counts stored in the PDS cube can be converted into physical units of spectral radiance Rad (W m-2 µm-1 sterad-1) by using the following equation:

S(λ(b), s, l) = Ns(b, s, l) / (ITF( λ(b), s)* ti)

where:

-S(λ(b),s,l) is the cube calibrated in spectral radiance which have the same bands and samples number of the raw data cube and a number of lines equal to the original minus the number of dark current frames of the raw cube;

- λ(b) is the wavelength associated to band b according to spectral calibration tables of VIS and IR channels (files DAWN_VIR_VIS_HIGHRES_SPECAL_Vx.TAB and DAWN_VIR_IR_HIGHRES_SPECAL_Vx.TAB, respectively);

-s, l corresponds to sample and line location of the pixel in the dark subtracted cube;

-ti is the integration time of the observations (in seconds) as indicated in the field FRAME_PARAMETER .EXPOSURE_DURATION of PDS header of the file for VIS and IR channels;

-ITF(λ(b), s) is the Instrument Transfer Function matrix for VIS and IR channels (files DAWN_VIR_VIS_RESP_Vx.DAT and DAWN_VIR_IR_RESP_Vx.DAT, respectively).

At the same time, to calculate the calibrated reflectance factor (sometimes termed 'I/F'), the equation is:

R(λ(b), s, l) = (S(λ(b), s, l) * ( π * (ssd / K) 2)) / si

where

-R(λ(b), s, l) is the cube calibrated reflectance factor which has the same number of bands and samples as the raw data cube and a number of lines equal to the original minus the number of dark current frames of the raw cube;

- K is the value of one astronomical unit expressed in km ( 149597870.7 ) ;

-ssd is the spacecraft heliocentric distance expressed in km, as read from the cube label file in the SPACECRAFT_SOLAR_DISTANCE field;

- si is the solar irradiance for VIS and IR channels (files DAWN_VIR_VIS_SOLAR_SPECTRUM_Vx.DAT and DAWN_VIR_IR_SOLAR_SPECTRUM_Vx.DAT, respectively).

These calculations can be applied to high resolution acquisitions (432 bands times 256 samples); in nominal modes, where spatial and/or spectral resolutions are reduced, it is necessary to interpolate both spectral tables and responsivity matrices according to binning values.

The following calibration files are stored in the CALIB directory of the PDS archives:

- DAWN_VIR_VIS_RESP_Vx.DAT, a 432x256 floating precision matrix containing the VIR-VIS Instrumental Transfer Function, including the VIS flat-Field. The file format is binary, matrix values are stored using double precision floating point precision, band interleaved (PDS type is IEEE_REAL with 8 bytes length)

- DAWN_VIR_IR_RESP_Vx.DAT, 432x256 floating precision matrix containing the VIR-IR Instrumental Transfer Function, including the IR flat-Field.

- DAWN_VIR_VIS_HIGHRES_SPECAL_Vx.TAB and

- DAWN_VIR_IR_HIGHRES_SPECAL_Vx.TAB, 432 row ASCII tables containing the wavelengths of the VIS and IR channels in High Resolution Mode.

- DAWN_VIR_VIS_WIDTH432_Vx.TAB and

- DAWN_VIR_IR_WIDTH432_Vx.TAB, 432 row ASCII tables containing the width of the VIS and IR channels in High Resolution Mode.

These files must be used for cubes collected in High Resolution Mode.

Cubes in Nominal Mode (x3 binning along bands) can be calibrated by using the following spectral calibration files:

- DAWN_VIR_VIS_NOMRES_SPECAL_Vx.TAB and

- DAWN_VIR_IR_NOMRES_SPECAL_Vx.TAB, 144 row ASCII tables containing the wavelengths of the VIS and IR channels in Low Resolution Mode.

- DAWN_VIR_VIS_WIDTH144_Vx.TAB and

- DAWN_VIR_IR_WIDTH144_Vx.TAB, 144 row ASCII tables containing the width of the VIS and IR channels in Low Resolution Mode ("x" is a digit representing the version number of the file). The first release is "V1".

VIR data included in this release are calibrated by using this basic pipeline. Further improvements, based on the use of the internal calibration sequences, may be included in future data releases. The ITF is also currently under improvement. Calibrated values in the spectral range [2.818µm - 3.272µm] are still under verification. These values have been put to null in the ITF only for the acquisitions performed during the mission campaigns VSH and VH2; during these two campaigns, a major external contamination is taken as responsible for the artifacts affecting the spectral range around 3 micrometers. The effects of the contamination on this spectral range have been avoided for acquisitions performed after these campaigns. This result was obtained by commanding the instrument into a given operative mode characterized during the analysis of this phenomenon. Such analysis come to the decision of minimizing the duration of the cryo-cooler activity, switching it off whenever compatible with the planned observation. Specifically, the cryo-cooler was always turned off at the end of each observation sequence.

The team is also preparing a specific algorithm to apply a correction for these two campaigns that will be released in the future.

10. KNOWN INSTRUMENT ARTIFACTS

VIR spectra are affected by residual systematic errors due to imperfect radiometric and spectral calibration that influence the quality of imaging spectrometer data. These include systematic deviations from spectrum of the target due to imperfect radiometric standards, by spectral miscalibration, and by systematic errors resulting from uncertainties in the solar reference spectrum.

The spectral images can show 'stripes' due to the slight deviation that exists between the input/output transfer function of each sample of the detector. These stripes are particularly evident when the signal is very low.

Visible channels with a wavelength > 0.95 microns cannot be used for scientific analysis, because of the straylight effect which currently has no correction. By excluding these wavelengths from the spectrum, the offset between the visible and the infrared channels in the range where they meet disappears. In some sporadic cases where this offset can still be observed, the instrument team recommends scaling the IR channel to the VIS channel, as the latter is in a good agreement with the Framing Camera dataset.

Defective pixel and filters boundaries should be not considered for any scientific analysis. The filter boundaries and defective pixels are listed in the tables below. Bands contiguous to the filter boundaries may be affected by straylight: the presence and intensity of straylight depends on the specific conditions at which each spectrum has been acquired. Therefore, the user shall check the presence of this effect and eventually discard the bands affected.

A discontinuity in the band pass filters between filters 3 and 4 results in a discontinuity in the transmissivity near 2.4 microns. When the data are corrected for variations in transmissivity a spike is introduced at this discontinuity (see fig. 8b in [3]).

VIR spectra are affected by a positive slope in the VIS-NIR range when compared to ground based spectra of the same target (Vesta and Ceres). Although the origin of this effect is not currently understood, we decided to re-normalize the VIR dataset to correct this effect. The correction is a scale factor computed as the ratio between a ground based reference and VIR spectrum of Ceres. For further details see [8].

The instrument team is working to resolve some of the issues described here. The team plans to release, in the near future, a new level of derived data products that will partially solve the problems above reported. The 'artifact removed' data set is planned for future release.

Dropouts that may be observed in some images are due to instrument saturation. The spectral range in which these dropouts occur varies depending upon exposure duration, target topography, and incidence  angle.         

In the following table, the filter boundaries positions are listed.

   ===========================================================================
    CHANNEL   SAMPLE INTERVAL   BAND INTERVAL   WAVELENGTH INTERVAL (nm)
     VIS         1 - 256           222-223         673.30398 - 675.19621
     IR          1 - 256           49 - 54        1474.79668 - 1522.09328
     IR          1 - 256          156 - 161       2486.94392 - 2534.24052
     IR          1 - 256          290 - 293       3754.4928 - 3782.87076
     IR          1 - 256           357-360        4388.26724 - 4416.6452
   ===========================================================================  

[ VIR DETECTOR FILTERS BOUNDARIES]

The defective pixels are listed below.

   ===========================================================================
    CHANNEL   SAMPLE   BAND INTERVAL   WAVELENGTH INTERVAL (nm)
     VIS        30         308                 836.03576
     VIS        31         308                 836.03576
     VIS        47         409                1027.15099
     VIS        48       187-188         607.07593-608.96816
     VIS        49          59                 364.87049
     VIS        54         137                 512.46443
     VIS        71         215                 660.05837
     VIS       100          78                 400.82286
     VIS       108         413                1034.71991
     VIS       109          19                 289.18129
     VIS       111          19                 289.18129
     VIS       114         424                1055.53444
     VIS       118         363                 940.10841
     VIS       126         410                1029.04322
     VIS       130         292                 805.76008
     VIS       136         271                 766.02325
     VIS       139         235                 697.90297
     VIS       147         222                 673.30398
     VIS       150          54                 355.40934
     VIS       150          59                 364.87049
     VIS       150          78                 400.82286
     VIS       160         372                 957.13848
     VIS       162        36-37           321.3492-323.24143
     VIS       162         248                 722.50196
     VIS       162         330                 877.66482
     VIS       163        36-37           321.3492-323.24143
     VIS       163         248                 722.50196
     VIS       163         330                 877.66482
     VIS       165          32                 313.78028
     VIS       166          32                 313.78028
     VIS       166         173                 580.58471
     VIS       168         232                 692.22628
     VIS       169         363                 940.10841
     VIS       172         189                 610.86039
     VIS       173          92                 427.31408
     VIS       175         228                 684.65736
     VIS       175       266-267         756.5621-758.45433
     VIS       176         152                 540.84788
     VIS       176         229                 686.54959
     VIS       177         155                 546.52457
     VIS       179         196                 624.106
     VIS       181         249                 724.39419
     VIS       183         354                 923.07834
     VIS       186         238                 703.57966
     VIS       186         387                 985.52193
     VIS       188         276                 775.4844
     VIS       188         352                 919.29388
     VIS       189         294                 809.54454
     VIS       189         352                 919.29388
     VIS       189         391                 993.09085
     VIS       189         413               1034.71991
     VIS       190         195                622.21377
     VIS       191         411               1030.93545
     VIS       194         358                930.64726
     VIS       196         266                756.5621
     VIS       196         362                938.21618
     VIS       199        23-24        296.75021-298.64244
     VIS       203         257                739.53203
     VIS       203         370                953.35402
     VIS       204         257                739.53203
     VIS       207         265                754.66987
     VIS       211         291                803.86785
     VIS       216         287                796.29893
     VIS       222         249                724.39419
     VIS       222         338                892.80266
     VIS       223       339-340       894.69489-896.58712
     VIS       225         274                771.69994
     VIS       227         103                448.12861
     VIS       229         248                722.50196
     VIS       234         306                832.2513
     VIS       234         424               1055.53444
     VIS       238         249                724.39419
     VIS       238         277                777.37663
     VIS       238       416-417       1040.3966-1042.28883
     VIS       239         405               1019.58207
     VIS       241        15-16         281.61237-283.5046
     VIS       241       386-387        983.6297-985.52193
     VIS       242        15-16         281.61237-283.5046
     VIS       242         364                942.00064
     VIS       245         128                495.43436
     VIS       248       304-305        828.46684-830.35907
     VIS       250         223                675.19621
     VIS       251         223                675.19621
     VIS       252         274                771.69994
     VIS       253         307                834.14353
     IR          8          86               1824.79152
     IR         12         148               2411.26936
     IR         16         327               4104.48764
     IR         20        39-43        1380.20348-1418.04076
     IR         21        39-42        1380.20348-1408.58144
     IR         22        40-42        1389.6628-1408.58144
     IR         27         374               4549.07568
     IR         35         218               3073.42176
     IR         45         337               4199.08084
     IR         51         212               3016.66584
     IR         52         280               3659.8996
     IR         56         430               5078.7976
     IR         74         121               2155.86772
     IR         79         185               2761.2642
     IR         79         190               2808.5608
     IR         82         190               2808.5608
     IR         84         188               2789.64216
     IR         86         182               2732.88624
     IR         86         200               2903.154
     IR         92          30               1295.0696
     IR         94         189               2799.10148
     IR         99          73               1701.82036
     IR        100          73               1701.82036
     IR        101       223-224       3120.71836-3130.17768
     IR        102          72               1692.36104
     IR        102         223               3120.71836
     IR        102         225               3139.637
     IR        103         223               3120.71836
     IR        111         304               3886.92328
     IR        112          28               1276.15096
     IR        121         193               2836.93876
     IR        122         172               2638.29304
     IR        128         149               2420.72868
     IR        128         187               2780.18284
     IR        130         195               2855.8574
     IR        132         182               2732.88624
     IR        136         344               4265.29608
     IR        138       383-384       4634.20956-4643.66888
     IR        140         202               2922.07264
     IR        142       341-342       4236.91812-4246.37744
     IR        143         343               4255.83676
     IR        144         343               4255.83676
     IR        145         343               4255.83676
     IR        146         342               4246.37744
     IR        146         344               4265.29608
     IR        148         108               2032.89656
     IR        149       169-170       2609.91508-2619.3744
     IR        155           1               1020.74932
     IR        156         1-9         1020.74932-1096.42388
     IR        156         196               2865.31672
     IR        157         1-15        1020.74932-1153.1798
     IR        157          25               1247.773
     IR        158         9-17        1096.42388-1172.09844
     IR        159        14-18        1143.72048-1181.55776
     IR        160        19-20        1191.01708-1200.4764
     IR        160        28-29        1276.15096-1285.61028
     IR        161          26               1257.23232
     IR        161        28-29        1276.15096-1285.61028
     IR        161          181              2723.42692
     IR        171        57-64        1550.47124-1616.68648
     IR        172        57-64        1550.47124-1616.68648
     IR        172         227               3158.55564
     IR        173        59-68        1569.38988-1654.52376
     IR        174        60-67        1578.8492-1645.06444
     IR        175        61-63        1588.30852-1607.22716
     IR        191       111-112       2061.27452-2070.73384
     IR        192       110-113       2051.8152-2080.19316
     IR        193       111-112       2061.27452-2070.73384
     IR        193       245-246       3328.8234-3338.28272
     IR        219         428               5059.87896
     IR        227         211               3007.20652
     IR        228          79               1758.57628
     IR        228         222               3111.25904
     IR        229         116               2108.57112
     IR        234         175               2666.671
     IR        235         175               2666.671
     IR        235         226               3149.09632
     IR        236         186               2770.72352
     IR        237         129               2231.54228
     IR        238          38               1370.74416
     IR        241         233               3215.31156
     IR        243         202               2922.07264
     IR        244         228               3168.01496
     IR        245       191-192       2818.02012-2827.47944
     IR        250         414               4927.44848
   ===========================================================================  

[ VIR DETECTOR DEFECTIVE PIXELS ]

Note that for the VIS detector, pixels located at these coordinates are to be considered both as filter boundaries and defective:

11. REFERENCES

[1] E. Ammannito, PhD dissertation, Università degli studi di Padova, Centro Interdipartimentale di Studi e Attività Spaziali (CISAS), 2008. Available on line at: http://paduaresearch.cab.unipd.it/760/1/tesi_online.pdf (in Italian)

[2] E. Ammannito, G. Filacchione, A. Coradini, F. Capaccioni, G. Piccioni, M.C. De Sanctis,M. Dami, A. Barbis, Rev. Sci. Instrum. 77, 093109 (2006)

[3] M.C. De Sanctis • A. Coradini • E. Ammannito • G. Filacchione • M.T. Capria • S. Fonte • G. Magni • A. Barbis • A. Bini • M. Dami • I. Ficai-Veltroni • G. Preti • VIR Team, 2011. The VIR Spectrometer. Space Sci Rev

DOI 10.1007/s11214-010-9668-5

[4] Melchiorri, R., Piccioni, G., Mazzoni, A., 2003. Review of Scientific Instruments, vol. 74, number 8, 3796-3801.

[5] G. Filacchione, PhD dissertation, Universita degli studi di Napoli Federico II, 2006. (in Italian)

[6] G. Filacchione, E. Ammannito, A. Coradini, F. Capaccioni, G. Piccioni, M.C. De Sanctis, M. Dami, A. Barbis, Rev. Sci. Instrum. 77, 103–106 (2006)

[7] M. C. De Sanctis , J.-Ph. Combe, E. Ammannito, E. Palomba, A. Longobardo, T. B. McCord, S. Marchi, F. Capaccioni, M. T. Capria, D. W. Mittlefehldt, C. M. Pieters, J. Sunshine, F. Tosi, F. Zambon, F. Carraro, S. Fonte, A. Frigeri, G. Magni, C. A. Raymond, C. T. Russell, and D. Turrini, Detection of widespread hydrated materials on Vesta by the VIR imaging spectrometer on board the Dawn Mission, The Astrophysical Journal Letters, 758:L36 (5pp), 2012 October 20.

[8] F.G. Carrozzo A. Raponi, M. C. De Sanctis, E. Ammannito, M. Giardino, E. D'Aversa, S. Fonte, and F. Tosi, Artifacts reduction in VIR/Dawn data, Rev. Sci. Instrum., 87, Issue 12 (2016)